Performance Tests
TDS5000B Series Specificationsand Performance Verification 2-39
n. Move the signal to the CH 3 input connector, and then changethe
TriggerSOURCE to CH3(TDS5034B, TDS5054B, TDS5034BE, or
TDS5104B).
o. Check — Amplitude of each trace other than CH 3 is 0.1 6 division or
less (discount tracewidth). Enter the amplitude in the test record.
p. Move the signal to the CH 4 input connector, and then changethe
TriggerSOURCE to CH4(TDS5034B, TDS5054B, TDS5054BE, or
TDS5104B).
q. Check — Amplitude of each trace other than CH 4 is 0.1 6 division or
less (discount tracewidth). Enter the amplitude in the test record.
3. Disconnect the hookup:Disconnect the cable from the generator output at
the input connector of the channel.
Time Base System ChecksThe following proceduresverify those characteristics that relate to the time base
system and arelisted as checked under Warranted Characteristics in the
Specifications section.
Equipment
required
Onet ime-mark generator (Item 12), or
Sine wave generator(It em 13)
One50 Ω, precision coaxial cable (It em 4)
Prerequisites ReadPrerequisites on page 2 --17and footnote warnings on page 2--19.
TDS5000BSeries oscilloscope
50Ωcoaxial cable
Output
TimeMark o r
SineWave
Generator
Figure2 -16: Initial test hookup
1. Install the test hookupand preset the instrument controls: