Performance Tests
2-56 TDS5000B Series Specifications and PerformanceVerification
Checkoutput
Figure2 -23: Measurement of main trigger out limits
3. Confirm SIGNAL OUT is within limits for gain:
a. Measuregain:
HMove the precision 50 cablefrom the AUX OUT BNC to the
SIGNAL OUT BNC.
HSet CH 2 Vertical SCALE to 100 mV.
HPressPUSH TO SET 50%.
HFrom the toolbar,click MEA S, and select the Ampl tab.
HClick the Pk-Pk button.
HClick Close.
b. Check against limits:
HFrom the toolbar,click the Vert button, then click the Termination
1Mbutton.
HClick Close.
HCHECK that the readout CH2 Pk-Pk is between200 mV and
300 mV,inclusive.
HEnter the voltage in the test record.
HFrom the toolbar,click the Vert button; then click the Termination
50 button.
HClick Close.