Performance Tests
2-30 TDS5000B Series Specifications and PerformanceVerification
HPress the front-panel Vertical button that corresponds to the channel
that you areto confirm.
HMove the leveling output of the sine wave generatorto the channel
that you selected.
b. Match the trigger sourceto the channel selected:
HPress the TriggerSOURCE button until the source that corresponds
to the channelthat you are to confirm is on.
c. Set its input impedance:
HFrom the toolbar,click the VERT button and select the tab for the
channel thatyou are to confirm. Click the Termination 50 button.
d. Set the verticalscale: Set the vertical SCALE that corresponds to the
channel thatyou are to confirm to one of the settings listed in Table 2--3
not yet checked.(Start with the 100 mV setting.)
e. Set the triggering coupling:
HClick the Coupling DC button.
Table2- 3: Analog bandwidth
Testfrequency
Vertical
scale
Reference
amplitude
(
6division)
Horizontal
scale
TDS5032B
TDS5034B
TDS5052B
TDS5054B
TDS50504BE TDS5104B -3 db Limits
1mV 6mV 1ns 150 MHz 175M Hz 175M Hz 4.24mV
2mV 12mV 1ns 250 MHz 300MHz 1GHz 8.48 mV
5mV 30mV 1ns 350 MHz 500MHz 1GHz 21.2 mV
10m V 60mV 1ns 350 MHz 500MHz 1GHz 42.4 mV
20m V 120mV 1ns 350MHz 500 MHz 1GHz 84.8 mV
50m V 300mV 1ns 350MHz 500 MHz 1GHz 212m V
100m V 600mV 1ns 350MHz 500MHz 1GHz 424 mV
200m V 1.2V 1ns 350MHz 500MHz 1GHz 848 mV
500m V 3V
11ns 350MHz 500MHz 1GHz 2.12 V1
1V 6V
11ns 350MHz 500MHz 1GHz 4.24 V1
1Ifyour generator cannot output 6 divisi ons of amplitude, determine its maximum output at the Test
frequency,anduse this for the referenceamplitude. The - 3 dblimit can be calculated as: 0.707×