Specifications
1-24 TDS5000B Series Specifications and PerformanceVerificationTable1- 11: Certifications and compliances (cont.)
Category Standardsor description
ECDecl aration of Conformity --
LowVoltage
Compliance was demonstrated to the following specification as listed in t he Official Journal of the
EuropeanUnion:
LowVoltage Directive 73/ 23/EEC, amended by 93/68/EEC
EN61010-1/A2: 1995 Safety requirements for electrical equipment for measurement
control and laboratory use.
U.S. Nationally Recognized
TestingLaboratory Listing
UL3111-1,First Edition Standardf orel ectrical measuring and test equipment.
Canadian Certification CAN/CSA C22.2, Safety requirements for electrical equipment for measurement,
No. 1010.1-92 control, and laboratoryuse.
Additional Compliance IEC61010-1/A2 Safet y requirements for electrical equipment for measurement,
control, and laboratory use.
Installation (Overvoltage)
Category
Terminalson this product may have different instal lation (overvoltage) category designations. The
installation categories are:
CATIII Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CATII Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools,and similar products. Equipment is usually cord-connected.
CATI Secondary(signal l evel) orbat tery operated circuits of electronic equipment.
Pollution Degree Am easureof the contaminates that could occur in the environment around and within a product.
Typicallythe i nternal environment inside a product is considered to be the same as the external.
Productsshoul d be used only in the environment for which they are rated.
Pollution Degree 2 Normally only dry,nonconductive poll ution occurs. Occasionally a
temporaryconduct ivity that is caused by condensation must be
expected. This location is a typical office/home environment.
Temporarycondensation occurs only when the product is out of
service.
SafetyCertification Compliance
Equipment Type Testand measuring
SafetyClass Class1 (as defined in I EC 61010-1/A2) --grounded product
Pollution Degree PollutionDegree 2 as defined in IEC 61010-1/A2
Rated for indoor use only