Performance Tests
TDS5000B Series Specificationsand Performance Verification 2-61
Sine Wave GeneratorLeveling Procedure
Some proceduresin this manual requirea leveled sine wave generator to produce
the necessarytest signals. If you do not have a leveled sine wavegenerator, use
one of the following proceduresto level the output amplitude of your sine wave
generator.
If you have a power divider,use this setup to monitor the generator output
simultaneously with the levelmeter and the oscilloscope.
Equipment
required
Sine wave generator(It em 13)
Level meter and power sensor (Item 14)
Powerdivi der (Item 15)
Twomale N to fem ale BNC adapters (Item 20)
Oneprecisi on coaxial cable (Item 4)
Prerequisites ReadPrerequisites on page 2 --17and footnote warnings on page 2--19.
1. Install the test hookup:Connect the equipment as shown in Figure 2 --28.
TDS5000BSeries oscilloscope
Sinewave
generator Levelmeter
Powersensor
Powerdivider
Attenuators
(ifnecessary)
Input
Output
Figure2 -28: Sine wave generator leveling equipment setup (Method A)

Method A