Brief Procedures
TDS5000B Series Specificationsand Performance Verification 2-13
TDS5000BSeries oscilloscope
Probefrom PROBE
COMPENSATION
outputto CH 1 input
Figure2 -6: Setup for the filesyst em test
3. Insert the testdisk: Insert the floppy disk in the floppy disk drive at the top
of the front panel. If your instrumentdoes not include a floppy disk drive,
see the Note on page 2--6.
4. Set up the oscilloscope: Push the front panel AUTOSET button.
5. Set the time base: Set the horizontal SCALE to 1ms/div. The time-base
readout is displayedat the bottom of the graticule.
6. Save the settings:
a. From the File menu select Save As. This displays the Save As dialog
box.
b. In the Save What field clickWaveform.
c. In the Sourcedrop-down list box select CH1.
d. In the Save in drop-downlist box select 31/2Floppy.
e. Note the default filename.
f. Click the Save button to save the waveformto the floppy disk.
7. Change the settings again: Set the horizontal SCALE to 200 s/div.
8. Verifythe file system works:
a. From the File menu select Recall. This displays the Recall dialo g box.
b. In the Recall What field click Waveform.
c. In the Look in drop-downlist box select 31/2Floppy. If your instrument
does not include a floppy disk drive, see the Note on page 2--6.
d. Locate andselect the waveform file name you previously stored.
e. Click the Recall button to display the stored wav eform on screen.