Performance Tests
2-36 TDS5000B Series Specifications and PerformanceVerification
HSelect Ref 2 from the Source drop-downlist box; then click the
Display Off button to toggle it to On. Youmay notice their
overlapping waveformhandle icons. See Figure 2--14 on page 2--36.
HClick the Close button.
f. Measurethe test signal:
HLocate the time referencepoints for these waveforms by first
identifying the point where the rising edgeof the left-most wave-
form crossesthe center horizontal graticule line, then note the
correspondingt ime referencepo int for the right-most waveform.See
Figure 2--14.
HPressCURSORS and select the V Bars Cursors Type; then clickthe
Close button.
HAlign one V bar cursor to the time reference point of the left-most
waveformedge and the other cursor to the time reference point of the
right-most waveformedge by rotating the multipurpose knobs (if
necessary,press the FINE buttons). See Figure 2--14.
HRead the measurement results at the :cursor readout on the screen.
Locatethe time reference
pointsfor these waveforms.
2
Displaythe waveforms. 1
Readresults.
4
Aligneach cursor to the time
referencepoints.
3
Figure2 -14: Measurement of channel delay