Performance Tests
2-42 TDS5000B Series Specifications and PerformanceVerification
This procedurechecks the sample rate portion of the Delta Time Measurement
Accuracy as listed in the Specifications section.The previous procedure, Check
Long-TermSample Rate and Delay Time Accuracy, on page 2--39,verified the
“PPM” portion of the delta time specification.
1. Install the test hookupand preset the instrument controls:
a. Initialize the oscilloscope:Press the DEFAULT SETUP button.
b. Hook up the pulse generator as shown in Figure 2--18.
TDS5000BSeries oscilloscope
Pulse
generator
50Ωcable
Output
20I50Ωcable
20I50Ωcable
SMAmale
tomale
SMA
short
SMAmale
tomale SMA
short
BNCT
connector
2Xattenuator
BNC90°female
tomale adapter
SMAT
connector
BNCto SMA
adapter
Figure2 -18: Delta time accuracy test hookup
HSet the pulse generatoroutput for a positive-going pulse with a
280 ps --800 ps rise-time, as shown in Table2--4 on page 2--44, and
for the fastest possible rep rate (at least 1 kHz).
HSet the pulse generatoroutput for about 500 mV. (This amplitude
can be adjustedlater to get a 5-division pulse on the screen.)
c. Modify the initialized front-panelcontrol settings:
HFrom the toolbar,click the Vert button. Set the termination of the
channel to 50 Ωby selectingthe channel tab and clicking the
Termination50 Ωbutton.
HPressAUTOSET. You may see both positive and negative pulses.
Adjust the TriggerLEVEL knob so that the trigger level is about
50% of the rising edge of the positive pulse.