Olicom 9600 manual Return Gas Coupling, Cooling Water OUT, Cryopump Electrical Outlet

Page 17

9600 Compressor Description

CTI-CRYOGENICS

HELIX TECHNOLOGY CORPORATION

Return Gas Coupling

The Return Gas Coupling returns the helium, which has been cycled through the cryopump, back to the Compressor. Refer to CTI-CRYOGENICS Helium Refrigeration System within this section for more information.

Cooling Water IN

The Cooling Water IN connector provides water to the Compressor from your facility to cool the Compressor during operation. The connector thread size is a 1/2 in. female pipe thread. The water must meet the specifications outlined in Table 1-3. Refer to Section 3 - Installation for more information on cooling water connections.

Cooling Water OUT

The Cooling Water OUT connector returns the water that has been used to cool the Compressor to your facility. The connector thread size is a

1/2 in. female pipe thread. Refer to Section 3 - Installation for more information on cooling water connections.

Cryopump Electrical Outlet

The Cryopump Electrical Outlet provide power to a single On-Board or Cryo-Torr Cryopump, an On-Board Splitter Box, or a Cryo-Torr Interface. The Compressor requires the use of an On-Board Splitter Box or Cryo-Torr Interface for multiple cryopump system connections. Refer to Table 1-5 for connector pin identification. Refer to Multiple On-Board Cryopump Connections or Multiple Cryo-Torr Cryopump Connections within this section for more information.

Table 1-5: Cryopump Electrical Outlet Pin Assignments

Identifier

Function

 

 

A and B

Heater Power - 208 VAC nominal

CCenter tap for D and E

D and E 24 VCT @ 4.6 Amps

F-G and G-H Cold Head Voltage Output 130-160 VAC @ 4.5 Amps

 

 

J

Chassis Ground

 

 

K

Not Used

1-10

P/N 8040540

Image 17
Contents 8040540 Rev 12/2004 Compressor Installation, Operation, Maintenance InstructionsHelix Page Table of Contents Operation InstallationMaintenance Figures Tables 9600 Compressor Description Compressor ConfigurationsGeneral BenefitsCTI-CRYOGENICS Helium Refrigeration System System Documentation9600 Compressor Dimensions SpecificationsCompressor Weight WeightElectrical Input Specifications ElectricalCooling Water Cooling Water SpecificationsGeneral General Compressor Operating Specifications9600 Compressor Rear View Component Locations Component DescriptionPower Inlet Power on IndicatorSystem Circuit Breaker Control Circuit BreakerCooling Water OUT Return Gas CouplingCryopump Electrical Outlet Cryopump Electrical Outlet Pin AssignmentsCompressor Remote Connector Cryopump Power Phase SwitchCompressor Remote Connector Pin Assignments Multiple On-Board Cryopump Connections9600 Compressor Connected to Multiple On-Board Cryopumps 9600 Compressor Connected to Multiple Cryo-Torr Cryopumps Multiple Cryo-Torr Cryopump ConnectionsCTI-CRYOGENICS Introduction Unpacking and InspectionShipping Carton Inspection Removal from Shipping CartonCompressor Inspection Shipping Carton ContentsCompressor Helium Static Pressure VerificationInstallation END Flexible Water Lines Supply and Return Water Line ConnectionsHard Water Lines Electrical Connections Power Cable Preparation9600 Compressor Circuit Breaker Terminals Cover Removed Proceed with Cryopump Power Phase Switch Setting Phase CheckConnecting Connecting/Disconnecting Helium Flex LinesCryopump Power Phase Switch Setting Disconnecting Single On-Board Cryopump Connections Single On-Board Cryopump ConnectionsSingle Cryo-Torr Cryopump Connections Single Cryo-Torr Cryopump InstallationHelium Line Connections Power Cable Connections Multiple Cryo-Torr Cryopump Connections CTI-CRYOGENICS Recommended Multiple Cryo-Torr Cryopump Installation Alternative Multiple Cryo-Torr Cryopump Installation Operation Adjusting System Helium PressureStatic Helium System Pressure Verification 9600 Compressor Helium Static ChargeCompressor Operation Replacement of Helium Circuit Components CTI-CRYOGENICS Suggested Maintenance Equipment MaintenanceScheduled Maintenance Disconnecting Self Sealing Couplings Adsorber ReplacementDate 9600 Compressor Helium Pressure Control Components Reducing Helium PressureAdding Helium Increasing Helium PressureCTI-CRYOGENICS Product Information Appendix a Customer Support InformationMail Page Figure B-19600 Compressor Flow Diagram Appendix B Flow DiagramAppendix B Flow Diagram Technical Inquiries Appendix C Troubleshooting ProceduresTroubleshooting the Compressor Compressor Remote Table C-1 Compressor Troubleshooting ProceduresAppendix a to contact CTI-CRYOGENICS Appendix D Schematic Figure D-19600 Motor Drive Compressor Schematic