9600 Compressor Installation, Operation, and Maintenance
CTI-CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Cryopump Power Phase Switch Setting
NOTE:
1.Inspect the
2.If a
3.If a
NOTE: All Cryopumps being connected to the 9600 Compressor must operate at the same phase.
Connecting/Disconnecting Helium Flex Lines
CAUTION
Make sure the helium flex lines are connected and disconnected from the 9600 Compressor using the following procedure and as shown in Figure
Connecting
1.Remove all dust plugs and caps from the Gas Supply and Return lines, and the Compressor and cryopump Supply and Return connectors. Check for the presence of a flat gasket in the male connector, and no gasket in the female connector.
2.Connect the Gas Return line to the GAS RETURN connector on the rear of the Compressor and then to the GAS RETURN connector on the cryopump. Using two wrenches as shown in Figure
3.Connect the Gas Supply line to the GAS SUPPLY connector on the rear of the Compressor and then to the GAS SUPPLY connector on the cryopump. Using two wrenches as shown in Figure
4.Attach the Supply and Return line identification labels to each end of the appropriate lines.
P/N 8040540 |