Appendix C - Troubleshooting Procedures
CTI-CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Table C-1: Compressor Troubleshooting Procedures
Problem |
| Possible Cause |
| Corrective Action |
|
|
|
|
|
1) System circuit breaker | 1) | Incorrect phasing of input | 1) | Check phasing of input |
(CB1) trips immediately |
| power. |
| power. Refer to Phase |
to the OFF (0) position |
|
|
| Check in Section 3. |
when switched to the ON |
|
|
|
|
(1) position. |
|
|
|
|
|
|
|
|
|
2) System (CB1) and Con- | 1) | No power coming from | 1) | Check source fuses, cir- |
trol Circuit (CB2) circuit |
| source. |
| cuit breakers, and wiring |
breakers remain in the |
|
|
| associated with the |
ON (1) position when |
|
|
| power source. Repair as |
switched ON but the |
|
|
| needed. |
compressor will not run. |
|
|
|
|
| 2) | Insufficient power | 2) | Verify adequate |
|
|
|
| phase input voltage. |
|
|
|
| Refer to Table |
| 3) | Remote control jumper | 3) | Check to insure that |
|
| plug not in place. This will |
| remote jumper plug is |
|
| apply only if remote cir- |
| fully seated. See Figure |
|
| cuit is not being used. |
| |
|
|
|
| Compressor Remote |
|
|
|
| Connector in Section 1 |
|
|
|
| for more information. |
| 4) | Improperly wired external | 4) | Verify correct installation |
|
| remote control circuit. |
| of remote control feature. |
|
| NOTE: Only applies if |
| Refer to Table |
|
| remote control feature is |
|
|
|
| being used. |
|
|
|
|
|
|
|
P/N 8040540 |