Olicom 9600 manual Reducing Helium Pressure, Compressor Helium Pressure Control Components

Page 47

Maintenance

CTI-CRYOGENICS

HELIX TECHNOLOGY CORPORATION

Adjusting System Helium Pressure

NOTE: These procedures can be performed on a compressor that is turned ON or OFF. However, the helium pressure gauge should be set to the static helium charge pressure value if the compressor is turned OFF or set to the normal system operating pressure if the compressor is turned ON. Refer to Section 4 - Operation for more information.

Reducing Helium Pressure

NOTE: You must obtain the normal system operating pressure from the Compressor Operation procedure in Section 4 - Operation in order to perform this procedure. If the normal system operating pressure is unknown, then shut the compressor OFF and perform the Static Helium System Pressure Verification procedure in Section 4 - Operation instead.

1.Remove the flare cap from the gas charge fitting as shown in Figure 5-3.

GAS CHARGE FLARED FITTING

HELIUM PRESSURE

GAUGE

GAS CHARGE

CONTROL VALVE

REAR PANEL

Figure 5-3: 9600 Compressor Helium Pressure Control Components

2.Open the gas charge control valve very slowly to allow a slight amount of helium to escape. Leave the valve open until the helium pressure gauge indicates one of the following:

To the appropriate value in Table 4-1 if the compressor is OFF and acclimated to a temperature between 60º F and 80º F

(15.5º C - 26.6º C).

5-4

P/N 8040540

Image 47
Contents 8040540 Rev 12/2004 Compressor Installation, Operation, Maintenance InstructionsHelix Page Table of Contents Operation InstallationMaintenance Figures Tables Benefits Compressor Configurations9600 Compressor Description GeneralCTI-CRYOGENICS Helium Refrigeration System System Documentation9600 Compressor Dimensions SpecificationsElectrical WeightCompressor Weight Electrical Input SpecificationsCooling Water Cooling Water SpecificationsGeneral General Compressor Operating Specifications9600 Compressor Rear View Component Locations Component DescriptionControl Circuit Breaker Power on IndicatorPower Inlet System Circuit BreakerCryopump Electrical Outlet Pin Assignments Return Gas CouplingCooling Water OUT Cryopump Electrical OutletMultiple On-Board Cryopump Connections Cryopump Power Phase SwitchCompressor Remote Connector Compressor Remote Connector Pin Assignments9600 Compressor Connected to Multiple On-Board Cryopumps 9600 Compressor Connected to Multiple Cryo-Torr Cryopumps Multiple Cryo-Torr Cryopump ConnectionsCTI-CRYOGENICS Removal from Shipping Carton Unpacking and InspectionIntroduction Shipping Carton InspectionHelium Static Pressure Verification Shipping Carton ContentsCompressor Inspection CompressorInstallation END Flexible Water Lines Supply and Return Water Line ConnectionsHard Water Lines Electrical Connections Power Cable Preparation9600 Compressor Circuit Breaker Terminals Cover Removed Proceed with Cryopump Power Phase Switch Setting Phase CheckConnecting Connecting/Disconnecting Helium Flex LinesCryopump Power Phase Switch Setting Disconnecting Single On-Board Cryopump Connections Single On-Board Cryopump ConnectionsSingle Cryo-Torr Cryopump Connections Single Cryo-Torr Cryopump InstallationHelium Line Connections Power Cable Connections Multiple Cryo-Torr Cryopump Connections CTI-CRYOGENICS Recommended Multiple Cryo-Torr Cryopump Installation Alternative Multiple Cryo-Torr Cryopump Installation 9600 Compressor Helium Static Charge Adjusting System Helium PressureOperation Static Helium System Pressure VerificationCompressor Operation Replacement of Helium Circuit Components CTI-CRYOGENICS Suggested Maintenance Equipment MaintenanceScheduled Maintenance Disconnecting Self Sealing Couplings Adsorber ReplacementDate 9600 Compressor Helium Pressure Control Components Reducing Helium PressureAdding Helium Increasing Helium PressureCTI-CRYOGENICS Product Information Appendix a Customer Support InformationMail Page Figure B-19600 Compressor Flow Diagram Appendix B Flow DiagramAppendix B Flow Diagram Technical Inquiries Appendix C Troubleshooting ProceduresTroubleshooting the Compressor Compressor Remote Table C-1 Compressor Troubleshooting ProceduresAppendix a to contact CTI-CRYOGENICS Appendix D Schematic Figure D-19600 Motor Drive Compressor Schematic