Electrical Connections

The following procedures provide information for making all three phase (180 - 250 VAC) electrical connections to the Compressor.

Power Cable Preparation

WARNING

Follow all local high voltage safety precautions when performing this procedure to reduce the possibility of electrical shock. Make sure all electrical power is OFF before proceeding with this procedure.

CAUTION

The cable used for making the Compressor power cable must be 10 gauge, 3 conductor cable with ground rated at 600 VAC.

1.Cut a 10 AWG (6.00 mm2), 3 conductor cable with ground to an appropriate length.

2.Strip the cable jacket back 4 in. (101.6 mm).

3.Strip the insulation back 3/8 in. (9.3 mm) on each individual conductor.

4.Install a #10 ring tongue terminal on the end of each conductor using the appropriate size double crimping tool.

5.Remove the rear panel as shown in Figure 3-2.

6.Remove the circuit breaker terminal cover as shown in Figure 3-2.

7.Install the cable into the Compressor through the cable strain relief.

8.Remove the 10-32 nut and install the grounding wire on the ground stud. Install the nut and tighten to 18 in.-lbs (0.21m-kg).

NOTE: Use a slotted screwdriver which is capable of holding a screw when performing steps 9 and 10.

9.Remove the screws from the Compressor circuit breaker terminals X, Y, and Z as shown in Figure 3-2.

3-4

P/N 8040540

Image 27
Contents 8040540 Rev 12/2004 Compressor Installation, Operation, Maintenance InstructionsHelix Page Table of Contents Installation MaintenanceOperation Figures Tables Benefits Compressor Configurations9600 Compressor Description GeneralCTI-CRYOGENICS Helium Refrigeration System System Documentation9600 Compressor Dimensions SpecificationsElectrical WeightCompressor Weight Electrical Input SpecificationsCooling Water Cooling Water SpecificationsGeneral General Compressor Operating Specifications9600 Compressor Rear View Component Locations Component DescriptionControl Circuit Breaker Power on IndicatorPower Inlet System Circuit BreakerCryopump Electrical Outlet Pin Assignments Return Gas CouplingCooling Water OUT Cryopump Electrical OutletMultiple On-Board Cryopump Connections Cryopump Power Phase SwitchCompressor Remote Connector Compressor Remote Connector Pin Assignments9600 Compressor Connected to Multiple On-Board Cryopumps 9600 Compressor Connected to Multiple Cryo-Torr Cryopumps Multiple Cryo-Torr Cryopump ConnectionsCTI-CRYOGENICS Removal from Shipping Carton Unpacking and InspectionIntroduction Shipping Carton InspectionHelium Static Pressure Verification Shipping Carton ContentsCompressor Inspection CompressorInstallation END Supply and Return Water Line Connections Hard Water LinesFlexible Water Lines Electrical Connections Power Cable Preparation9600 Compressor Circuit Breaker Terminals Cover Removed Proceed with Cryopump Power Phase Switch Setting Phase CheckConnecting/Disconnecting Helium Flex Lines Cryopump Power Phase Switch SettingConnecting Disconnecting Single On-Board Cryopump Connections Single On-Board Cryopump ConnectionsSingle Cryo-Torr Cryopump Connections Single Cryo-Torr Cryopump InstallationHelium Line Connections Power Cable Connections Multiple Cryo-Torr Cryopump Connections CTI-CRYOGENICS Recommended Multiple Cryo-Torr Cryopump Installation Alternative Multiple Cryo-Torr Cryopump Installation 9600 Compressor Helium Static Charge Adjusting System Helium PressureOperation Static Helium System Pressure VerificationCompressor Operation Replacement of Helium Circuit Components CTI-CRYOGENICS Maintenance Scheduled MaintenanceSuggested Maintenance Equipment Disconnecting Self Sealing Couplings Adsorber ReplacementDate 9600 Compressor Helium Pressure Control Components Reducing Helium PressureAdding Helium Increasing Helium PressureCTI-CRYOGENICS Appendix a Customer Support Information MailProduct Information Page Figure B-19600 Compressor Flow Diagram Appendix B Flow DiagramAppendix B Flow Diagram Appendix C Troubleshooting Procedures Troubleshooting the CompressorTechnical Inquiries Compressor Remote Table C-1 Compressor Troubleshooting ProceduresAppendix a to contact CTI-CRYOGENICS Appendix D Schematic Figure D-19600 Motor Drive Compressor Schematic