9600 Compressor Installation, Operation, and Maintenance

CTI-CRYOGENICS

HELIX TECHNOLOGY CORPORATION

CRYOPUMP MOTOR

PHASE SWITCH

CIRCUIT BREAKER TERMINAL COVER (SEE DETAIL A)

REAR PANEL

CRYOPUMP

PHASE

X Y Z

THREE PHASE POWER

DETAIL A (COVER REMOVED)

Figure 3-2: 9600 Compressor Circuit Breaker Terminals (Cover Removed)

NOTE: The phase order in which the conductor terminal lugs are connected to circuit breaker terminals X, Y, and Z will be determined during the Phase Check Procedure.

NOTE: For installation where one of the three phase legs is at or near ground potential, connect that leg to terminal Y on the Compressor as shown in Figure 3-2.

10.Install the conductor terminal lugs to the circuit breaker terminals X, Y, and Z as shown in Figure 3-2.

11.Torque the screws to 12 in.-lbs (0.14m-kg).

12.Allow enough cable to stay in the electrical enclosure to prevent strain on the electrical connections and tighten the screws on the cable strain relief.

13.Install the power source end of the power cable according to the local electrical codes.

14.Install the circuit breaker terminal cover.

15.Proceed with Phase Check.

P/N 8040540

3-5

Image 28
Contents Compressor Installation, Operation, Maintenance Instructions 8040540 Rev 12/2004Helix Page Table of Contents Maintenance InstallationOperation Figures Tables Compressor Configurations 9600 Compressor DescriptionGeneral BenefitsSystem Documentation CTI-CRYOGENICS Helium Refrigeration System9600 Compressor Specifications DimensionsWeight Compressor WeightElectrical Input Specifications ElectricalCooling Water Specifications Cooling WaterGeneral Compressor Operating Specifications GeneralComponent Description 9600 Compressor Rear View Component LocationsPower on Indicator Power InletSystem Circuit Breaker Control Circuit BreakerReturn Gas Coupling Cooling Water OUTCryopump Electrical Outlet Cryopump Electrical Outlet Pin AssignmentsCryopump Power Phase Switch Compressor Remote ConnectorCompressor Remote Connector Pin Assignments Multiple On-Board Cryopump Connections9600 Compressor Connected to Multiple On-Board Cryopumps Multiple Cryo-Torr Cryopump Connections 9600 Compressor Connected to Multiple Cryo-Torr CryopumpsCTI-CRYOGENICS Unpacking and Inspection IntroductionShipping Carton Inspection Removal from Shipping CartonShipping Carton Contents Compressor InspectionCompressor Helium Static Pressure VerificationInstallation END Hard Water Lines Supply and Return Water Line ConnectionsFlexible Water Lines Power Cable Preparation Electrical Connections9600 Compressor Circuit Breaker Terminals Cover Removed Phase Check Proceed with Cryopump Power Phase Switch SettingCryopump Power Phase Switch Setting Connecting/Disconnecting Helium Flex LinesConnecting Disconnecting Single On-Board Cryopump Connections Single On-Board Cryopump ConnectionsSingle Cryo-Torr Cryopump Installation Single Cryo-Torr Cryopump ConnectionsHelium Line Connections Power Cable Connections Multiple Cryo-Torr Cryopump Connections CTI-CRYOGENICS Recommended Multiple Cryo-Torr Cryopump Installation Alternative Multiple Cryo-Torr Cryopump Installation Adjusting System Helium Pressure OperationStatic Helium System Pressure Verification 9600 Compressor Helium Static ChargeCompressor Operation Replacement of Helium Circuit Components CTI-CRYOGENICS Scheduled Maintenance MaintenanceSuggested Maintenance Equipment Adsorber Replacement Disconnecting Self Sealing CouplingsDate Reducing Helium Pressure 9600 Compressor Helium Pressure Control ComponentsIncreasing Helium Pressure Adding HeliumCTI-CRYOGENICS Mail Appendix a Customer Support InformationProduct Information Page Appendix B Flow Diagram Figure B-19600 Compressor Flow DiagramAppendix B Flow Diagram Troubleshooting the Compressor Appendix C Troubleshooting ProceduresTechnical Inquiries Table C-1 Compressor Troubleshooting Procedures Compressor RemoteAppendix a to contact CTI-CRYOGENICS Appendix D Schematic Figure D-19600 Motor Drive Compressor Schematic