9600 Compressor Installation, Operation, and Maintenance
CTI-CRYOGENICS
|
|
|
|
|
|
|
| HELIX TECHNOLOGY CORPORATION | |
|
| Table | |||||||
|
|
|
|
|
|
|
| ||
|
| Problem |
| Possible Cause |
| Corrective Action |
| ||
|
|
|
|
|
|
| |||
| 3) System circuit breaker | 1) | Damaged | 1) | Check for compressor |
| |||
|
| (CB1) will not remain in |
| power cable, connectors, |
| operation with cryopump |
| ||
|
| the ON (1) position |
| or drive motor. |
| cable disconnected from |
| ||
|
| when switched ON. The |
|
|
| compressor. Refer to |
| ||
|
| Control Circuit circuit |
|
|
| Appendix A to contact |
| ||
|
| breaker (CB2) trips |
|
|
| the Customer Support |
| ||
|
| when excessive current |
|
|
| Center if the compressor |
| ||
|
| is being drawn by the |
|
|
| operates improperly. |
| ||
|
| cold head or 24 volt |
|
|
|
|
|
|
|
|
| compressor control cir- | 2) | Damaged component in | 2) | Refer to Appendix A to |
| ||
|
| cuits. |
| the compressor power or |
| contact the Customer |
| ||
|
|
|
| control circuit. |
| Support Center. |
| ||
|
|
|
|
|
|
|
| ||
| 4) | System circuit breaker | 1) | Thermal protective | 1) | Check for inadequate |
| ||
|
| (CB1) remains in the ON |
| switches are open. |
| water cooling. Refer to |
| ||
|
| (1) position and the com- |
|
|
| Table |
| ||
|
| pressor stops after sev- |
|
|
|
|
|
|
|
|
| eral minutes of operation | 2) Very cold water has caused | 2) | Recheck for proper cool- |
| |||
|
| and remains OFF (0). |
| a restriction of oil flow |
| ing water temperature. |
| ||
|
|
|
| through the oil injection |
| Refer to Table |
| ||
|
|
|
| orifice during |
| Restart compressor |
| ||
|
|
|
|
|
| repeatedly until continu- |
| ||
|
|
|
|
|
| ous operation is achieved. |
| ||
|
|
|
|
|
|
|
| ||
| 5) | System circuit breaker | 1) | Loss or degradation of | 1) | Check that line voltage is |
| ||
|
| (CB1) trips after a period |
| power from the source. |
| correct on all phases. |
| ||
|
| of running. |
|
|
|
|
|
|
|
|
|
| 2) | Defective motor windings. | 2) | Check running current on |
| ||
|
|
|
|
|
| all phases. |
| ||
|
|
|
|
| 3) | Refer to Appendix A to |
| ||
|
|
|
|
|
| contact the local Cus- |
| ||
|
|
|
|
|
| tomer Support Center. |
| ||
|
|
|
|
|
| ||||
| 6) All cryopumps making | 1) Cryopump power phase | 1) | Set power phase switch in |
| ||||
|
| loud noise. |
| switch is in the wrong |
| opposite position. |
| ||
|
|
|
| position. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
P/N 8040540 |