9600 Compressor Installation, Operation, and Maintenance
CTI-CRYOGENICS
HELIX TECHNOLOGY CORPORATION
Section 4 - Operation
Adjusting System Helium Pressure
Your
NOTE: The 9600 Compressor is designed for continuous operation and should remain ON even when the cryopumps are in a regeneration cycle.
Static Helium System Pressure Verification
The proper system static helium charge pressure is necessary so that the cryopumps operate at maximum performance as well as to assure that the compressor will operate below the maximum design motor winding temperature which will maximize the life of the compressor motor.
1.Make sure the Compressor and Cryopump(s) are OFF.
2.Make sure all system components are connected together as described in Section 3 - Installation.
3.Allow all system components to acclimate to a temperature between 60º F and 80º F (15.5º C - 26.6º C).
4.Read the compressor helium pressure gauge located on the com- pressor rear panel as shown in Figure
Table 4-1: 9600 Compressor Helium Static Charge
Line Frequency | Helium Static Charge Pressure |
|
|
60 Hz | 240 - 250 psig (16.5 - 17.2 bars ) |
|
|
50 Hz | 255 - 265 psig ( 17.6 |
|
|