Local Terminal Configuration

Using the Main Menu

4.Press a number to select a field and then type the corresponding values necessary to configure the probe to operate on your network.

Serial port IP address

Press 3 and then type the serial port IP address for the

 

probe. The default Serial Port IP Address is 0.0.0.0.

Serial port

Press 4 and then type the serial port subnet mask for

subnet mask

the probe (optional). It is recommended that you do not

 

change the serial port subnet mask unless there is a

 

conflict. The default Serial Port Subnet Mask is

 

255.255.255.192.

Serial port speed

Press 5 and then type a serial port speed (300 to 38,400

 

baud) for the probe’s SNMP connection. The default is

 

9600 baud. This speed is used only for Out-of-Band

 

access to the probe using SNMP. It does not affect the

 

serial connection for the local terminal, which is fixed at

 

9600 baud.

 

Make sure that the serial port speed is set to less than or

 

equal to the maximum speed of the modem to be used.

Serial port mode

Press 6 and then select the serial port mode by pressing 1

 

for direct connection (the default) or 2 for modem

 

connection.

Serial port hardware

Press 7 and then select hardware flow control Off by

flow control

pressing 1 or hardware flow control On (the default) by

 

pressing 2.

Modem Init String

Press 8 and then type the modem initialization string.

 

Only the first 20 characters of the 256 character

 

maximum will be displayed in the Modify/View Serial

 

Port Settings menu. The default is

 

^s^M^d1^sATEOQOV1X4 S0=1 S2=43^M.