Troubleshooting

In Auto mode, PRIMARY_VALUE reflects the value and status quality of the output channels.

Methods

Sensor Calibration

In order to calibrate the sensor, the following steps are performed by the user calibration method:

1.Set MODE_BLK.TARGET = OOS.

2.Apply desired pressure (low pressure); allow to stabilize. Pressure applied must be between range limits defined in

PRIMARY_VALUE_RANGE.

3.Set CAL_POINT_LO to applied pressure.

4.Apply desired pressure (high pressure); allow to stabilize. Pressure applied must be between range limits defined in PRIMARY_VALUE_RANGE and greater than CAL_POINT_LO + CAL_MIN_SPAN.

5.Set CAL_POINT_HI to applied pressure.

6.Set SENSOR_CAL_DATE to current date.

7.Set SENSOR_CAL_WHO to person responsible for calibration.

8.Set SENSOR_CAL_LOC to calibration location.

9.Set MODE_BLK.TARGET = AUTO.

Refer to the table below to troubleshoot any problems encountered.

Symptom

Possible Causes

Corrective Action

 

 

 

Mode will not leave

Target mode not set.

Set target mode to something other

OOS

 

than OOS.

 

 

 

 

Detailed status error

See “Diagnostics” on page C-5

 

Resource block

The actual mode of the Resource block

 

 

is OOS. See Resource Block

 

 

Diagnostics for corrective action.

 

 

 

Pressure or Sensor

Measurement or

See “Diagnostics” on page C-5

Temperature Status

Device Error

 

is BAD

 

 

 

 

 

RESOURCE BLOCK

Overview

This section contains information on the Rosemount 2051 Resource Block.

 

Descriptions of all Resource Block Parameters, errors, and diagnostics are

 

included. Also the modes, alarm detection, status handling, and

 

troubleshooting are discussed.

 

Definition

 

The resource block defines the physical resources of the device. The

 

resource block also handles functionality that is common across multiple

 

blocks. The block has no linkable inputs or outputs and it performs memory

 

diagnostics.

C-6

Page 136
Image 136
Emerson Process Management 2051 manual Troubleshooting, Overview, Rosemount, Reference Manual, Resource Block