RESOURCE BLOCK

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

RESOURCE BLOCK

This section describes error conditions found in the Resource block. Read

 

Table 5-2 through Table 5-4 to determine the appropriate corrective action.

Table 5-2. Resource Block

Block Errors

BLOCK_ERR messages

Table 5-2 lists conditions reported in the BLOCK_ERR parameter.

 

Table 5-3. Resource Block SUMMARY_STATUS messages

Condition Name and Description

Other

Simulate Active: This indicates that the simulation switch is in place. This is not an indication that the I/O blocks are using simulated data.

Device Fault State Set

Device Needs Maintenance Soon

Memory Failure: A memory failure has occurred in FLASH, RAM, or EEPROM memory

Lost Static Data: Static data that is stored in non-volatile memory has been lost.

Lost NV Data: Non-volatile data that is stored in non-volatile memory has been lost. Device Needs Maintenance Now

Out of Service: The actual mode is out of service.

Condition Name

Uninitilized

No repair needed

Repairable

Call Service Center

Table 5-4. Resource Block DETAILED_STATUS with recommended action messages

Condition Name

Recommended Action

 

 

 

LOI Transducer block error

1.

Restart processor

 

2.

Check display connection

 

3.

Call service center

 

 

 

Sensor Transducer block error.

1.

Restart processor

 

2.

Check sensor module cable

 

3.

Call service center

Mfg. Block integrity error

1.

Restart processor

 

2.

Call service center

 

 

 

Non-Volatile memory integrity error

1.

Restart processor

 

2.Call service center

ROM integrity error

1.

Restart processor

 

2.

Call service center

5-5