Emerson Process Management 2051 manual Rosemount, Reference Manual, Rem_Seal_Fill

Models: 2051

1 152
Download 152 pages 41.56 Kb
Page 139
Image 139
REM_SEAL_FILL

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

 

Index

 

 

Parameter

Number

Description

 

 

 

self_test

59

Instructs resource block to perform self-test. Tests are device specific.

distributor

42

Reserved for use as distributor ID. No Foundation enumerations defined at this time.

 

 

 

REM_SEAL_FILL

73

Type of fill fluid used in the remote seals.

 

 

2

= Silicone

 

 

3

= Syltherm 800

 

 

4

= Inert (Halocarbon)

 

 

5

= Glycerin and Water

 

 

6

= Propylene Glycol and Water

 

 

7

= Neobee M–20

 

 

251 = None

 

 

252 = "Unknown"

 

 

253 = "Special"

REM_SEAL_ISO_MAT

70

Type of material of the remote seal isolators. See remote seal number codes.

 

 

2

= 316L Stainless Steel

 

 

3

= Hastelloy C–276

 

 

5

= Tantalum

 

 

9

= Co–Cr–Ni

 

 

251 = None

 

 

252 = "Unknown"

 

 

253 = "Special"

 

 

 

REM_SEAL_NUM

71

Number of remote seals.

 

 

1

= One seal

 

 

2

= Two seals

 

 

251 = None

 

 

252 = "Unknown"

 

 

253 = "Special"

REM_SEAL_TYPE

66

Type of remote seals.

 

 

0

= Undefined

 

 

1

= Reserved

 

 

2

= CTW

 

 

3

= EFW (Expanded Flange Seal)

 

 

4

= PFW (Pancake)

 

 

5

= RFW (Flanged Remote)

 

 

6

= RTW (Threaded Remote)

 

 

7

= SCW

 

 

8

= SSW

 

 

9

= High Temperature

 

 

10 = FFW (Flanged Flush Surface)

 

 

11 = UCW

 

 

12 = TSW

 

 

251 = None

 

 

252 = "Unknown"

 

 

253 = "Special"

 

 

 

RESTART

16

Allows a manual restart to be initiated. Several degrees of restart are possible. They are

 

 

the following:

 

 

1

Run – nominal state when not restarting

 

 

2

Restart resource – not used

 

 

3

Restart with defaults – set parameters to default values. See

 

 

START_WITH_DEFAULTS below for which parameters are set.

 

 

4

Restart processor – does a warm start of CPU.

RS_STATE

07

State of the function block application state machine.

 

 

 

save_config_blocks

62

Number of EEPROM blocks that have been modified since last burn. This value will count

 

 

down to zero when the configuration is saved.

save_config_now

61

Controls saving of configuration.

 

 

 

security_IO

65

Status of security jumper/switch.

SHED_RCAS

26

Time duration at which to give up on computer writes to function block RCas locations.

 

 

Shed from RCas shall never happen when SHED_ROUT = 0

 

 

 

SHED_ROUT

27

Time duration at which to give up on computer writes to function block ROut locations.

 

 

Shed from ROut shall never happen when SHED_ROUT = 0

 

 

 

 

C-9

Page 139
Image 139
Emerson Process Management 2051 manual Rosemount, Reference Manual, Rem_Seal_Fill