Appendix A

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

Appendix A

Reference Data

PERFORMANCE SPECIFICATIONS

Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-1 Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-4 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-9 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-12 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-22 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-25

For zero based spans, reference conditions, silicone oil fill, SST materials, Coplanar flange (2051C) or 1/2 in. - 14 NPT (2051T) process connections, digital trim values range points. Applies to 4-20 mA HART output only unless otherwise noted.

Conformance To Specification (±3σ (Sigma))

Reference Accuracy(1)

Technology leadership, advanced manufacturing techniques and statistical process control ensure specification conformance to at least ±3σ.

Models(1)

 

Standard

 

 

 

 

 

Performance Option, P8

 

 

 

 

 

 

2051C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ranges 2-5 ±0.075% of span

 

 

 

 

Ranges 2-5

High Accuracy Option, P8

 

 

 

 

For spans less than 10:1, accuracy =

±0.065% of span

 

 

 

 

 

 

 

 

 

 

 

 

⎛ URL

 

 

For spans less than 10:1, accuracy =

 

±

 

0.025 + 0.005

 

% of Span

 

 

 

 

 

 

 

 

 

 

 

 

--------------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span

 

 

±

 

0.015 + 0.005

URL

 

% of Span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

--------------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span

 

 

 

Range 1 ±0.10% of span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For spans less than 15:1, accuracy =

 

 

 

 

 

 

 

 

 

 

±

 

0.025 + 0.005

⎛ URL

 

% of Span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

--------------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2051T

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ranges 1-4

±0.075% of span

 

 

 

 

Ranges 1-4 High Accuracy Option, P8

 

 

 

 

 

For spans less than 10:1, accuracy =

±0.065% of span

 

 

 

 

 

 

 

 

 

 

 

⎛ URL ⎞

 

 

 

 

For spans less than 10:1, accuracy =

 

 

±

0.0075

% of Span

 

 

 

 

 

 

 

 

 

 

 

--------------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span

 

 

 

 

±

0.0075

⎛ URL

% of Span

 

 

 

 

 

 

 

 

 

 

 

 

--------------

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Span

 

 

 

 

 

Range 5

±0.075% of span for spans greater than 5:1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2051L

Ranges 2-4 ±0.075% of span

For spans less than 10:1, accuracy =

±0.025 + 0.005--------------URL% of Span Span

(1)For FOUNDATION fieldbus transmitters, use calibrated range in place of span.

www.rosemount.com

Page 73
Image 73
Emerson Process Management 2051 Appendix A, Reference Data, Performance Specifications, Reference Accuracy1, Rosemount