Manual background

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

Flange Bolts

The 2051 is shipped with a Coplanar flange installed with four 1.75-in. (44

mm)flange bolts. See Figure 2-6 and Figure 2-7 on pages 2-14 and 2-14. Stainless steel bolts are coated with a lubricant to ease installation. Carbon steel bolts do not require lubrication. No additional lubricant should be applied when installing either type of bolt. Bolts are identified by their head markings:

B7M

316

B8M

F593_*

Carbon Steel (CS) Head Markings

Stainless Steel (SST) Head Markings

*The last digit in the F593_ head marking may be any letter between A and M.

Bolt Installation

Only use bolts supplied with the 2051 or provided by Emerson Process Management as spare parts. When installing the transmitter to one of the optional mounting brackets, torque the bolts to 125 in-lb. (0,9 N-m). Use the following bolt installation procedure:

1.Finger-tighten the bolts.

2.Torque the bolts to the initial torque value using a crossing pattern.

3.Torque the bolts to the final torque value using the same crossing pattern.

Torque values for the flange and manifold adapter bolts are as follows:

Table 2-3. Bolt Installation

Torque Values

Bolt Material

Initial Torque

Value

Final Torque Value

 

CS-ASTM-A449 Standard

300 in.-lb (34 N-m)

650 in.-lb (73 N-m)

 

 

 

 

 

 

316 SST—Option L4

150 in.-lb (17

N-m)

300 in.-lb (34 N-m)

 

 

 

 

 

 

ASTM-A-193-B7M—Option L5

300 in.-lb (34

N-m)

650 in.-lb (73 N-m)

 

 

 

 

 

 

ASTM-A-193 Class 2, Grade

150 in.-lb (17

N-m)

300 in.-lb (34 N-m)

 

B8M—Option L8

 

 

 

 

 

 

 

 

See “Safety Messages” on page 2-1 for complete warning information.

2-13

Page 23
Image 23
Emerson Process Management 2051 manual Rosemount, Reference Manual, 00809-0200-4101,Rev AA July