Ambient Temperature Effect per 50°F (28°C)

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

Ambient Temperature Effect per 50°F (28°C)

Models

Ambient Temperature Effect

 

 

2051C

Ranges 2-5 ±(0.025% URL + 0.125% span) from 1:1 to 5:1 ±(0.05% URL + 0.25% span) from 5:1 to 100:1

Range 1 ±(0.2% URL + 0.5% span) from 1:1 to 50:1

2051T

Range 2-4 ±(0.05% URL + 0.25% span) from 1:1 to 30:1 ±(0.07% URL + 0.25% span) from 30:1 to 100:1

Range 1 ±(0.05% URL + 0.25% span) from 1:1 to 10:1 ±(0.10% URL + 0.25% span) from 10:1 to 100:1

Range 5 ±(0.2% URL + 0.3% span)

2051LSee Instrument Toolkit

Mounting Position

Effects

Models

Mounting Position Effects

 

 

2051C

Zero shifts up to ±1.25 inH2O (3,1 mbar), which can be calibrated out. No span effect.

2051T

Zero shifts up to ±2.5 inH2O (6,2 mbar), which can be calibrated out. No span effect.

2051L

With liquid level diaphragm in vertical plane, zero shift of up to 1 inH2O (2,49 mbar). With diaphragm in

 

horizontal plane, zero shift of up to 5 inH2O (12,43 mbar) plus extension length on extended units. Zero

 

shifts can be calibrated out. No span effect.

Vibration Effect

Power Supply Effect

Electromagnetic Compatibility (EMC)

Transient Protection (Option Code T1)

Less than ±0.1% of URL when tested per the requirements of IEC60770-1 field or pipeline with high vibration level (10-60 Hz 0.21mm displacement peak amplitude / 60-2000 Hz 3g).

Less than ±0.005% of calibrated span per volt.

Meets all relevant requirements of EN 61326 and NAMUR NE-21.

Meets IEEE C62.41, Category Location B 6 kV crest (0.5 μs - 100 kHz)

3 kV crest (8 × 20 microseconds)

6 kV crest (1.2 × 50 microseconds)