Emerson Process Management 2051 manual Rosemount, Reference Manual, Mode will not leave

Models: 2051

1 152
Download 152 pages 41.56 Kb
Page 143
Image 143
Manual background

Reference Manual

00809-0200-4101, Rev AA July 2008

Rosemount 2051

Symptom

Possible Causes

Corrective Action

 

 

 

Mode will not leave

Target mode not set.

Set target mode to something other than

OOS

 

OOS.

 

 

 

 

Memory Failure

BLOCK_ERR will show the memory failure.

 

 

Check RESTART vallue. Restart the device

 

 

by setting RESTART to Processor. If the

 

 

block error does not clear, call the factory.

Block Alarms Will

Features

FEATURES_SEL does not have Reports

not work

 

enabled. Enable the Reports bit.

 

 

 

 

Notification

LIM_NOTIFY is not high enough. Set equal

 

 

to MAX_NOTIFY.

 

 

 

C-13