SWAVeform Subsystem

RANGe

Example

OUTPUT XXX;":MACHINE1:SWAVEFORM:MLENGTH 262144"

 

 

Query:MACHine{12}:SWAVeform:MLENgth?

The MLENgth query returns the current analyzer memory depth selection.

Returned Format [:MACHine{12}:SWAVeform:MLENgth] <memory_length><NL>

Example

OUTPUT XXX;":MACHINE1:SWAVEFORM:MLENGTH?"

 

 

RANGe

CommandMACHine{12}:SWAVeform:RANGe <number_of_samples>

<number_of_ samples>

Example

Query

Returned Format

Example

The RANGe command allows you to specify the number of samples across the screen on the State Waveform display. It is equivalent to ten times the states per division setting (states/Div) on the front panel. A number between 10 and 5000 may be entered.

integer from 10 to 5000

OUTPUT XXX;":MACHINE2:SWAVEFORM:RANGE 80"

MACHine{12}:SWAVeform:RANGe?

The RANGe query returns the current range value.

[MACHine{12}:SWAVeform:RANGe] <number_of_samples><NL>

OUTPUT XXX;":MACHINE2:SWAVEFORM:RANGE?"