SCHart Subsystem

CENTer

CENTer

CommandMACHine{12}:SCHart:CENTer <marker_type>

The CENTer command centers the chart display about the specified markers. The markers are placed in the SLISt subsystem.

<marker_type> {X O XO TRIGger}

Example

OUTPUT XXX;":MACHINE1:SCHART:CENTER XO"

 

 

 

HAXis

Command

MACHine{12}:SCHart:HAXis

 

{STAtes,<state_low_value>,<state_high_value>

 

<label_name>,<label_low_value>,<label_high_value>,

 

<state_low_value>,<state_high_value>}

 

The HAXis command allows you to select whether states or a label’s values

 

will be plotted on the horizontal axis of the chart. The axis is scaled by

 

specifying the high and low values. The shortform for STATES is STA. This

 

is an intentional deviation from the normal truncation rule.

<state_low_

integer from –516096 to +516096 (HP 16554A), or from –1040384 to

value>

+1040384 (HP 16555A), or from –2080768 to +2080768 (HP 16555D)

<state_high_

integer from <state_low_value> to +516096 (HP 16554A), +1040384

value>

(HP 16555A), or +2080768 (HP 16555D)

<label_name>

a string of up to 6 alphanumeric characters

<label_low_

string from 0 to 2321 (#HFFFFFFFF)

value>

 

<label_high_

string from <label_low_value> to 232–1 (#HFFFFFFFF)

value>

 

9–5