Image 184

 

 

TTRigger (TTRace) Subsystem

 

 

BRANch

 

 

Within the limitations shown by the syntax definitions, complex expressions

 

 

may be formed using the AND and OR operators. Expressions are limited to

 

 

what you could manually enter through the Timing Trigger menu. Regarding

 

 

parentheses, the syntax definitions on the next page show only the required

 

 

ones. Additional parentheses are allowed as long as the meaning of the

 

 

expression is not changed. Figure 12-2, on page 12-11, shows a complex

 

 

expression as seen in the Timing Trigger menu.

 

 

 

Example

The following statements are all correct and have the same meaning. Notice

 

 

that the conventional rules for precedence are not followed. The expressions

 

 

are evaluated from left to right.

 

 

OUTPUT XXX;":MACHINE1:TTRIGGER:BRANCH1 ’C AND D OR F OR G’, 1"

 

 

OUTPUT XXX;":MACHINE1:TTRIGGER:BRANCH1 ’((C AND D) OR (F OR

 

 

G))’, 1"

 

 

OUTPUT XXX;":MACHINE1:TTRIGGER:BRANCH1 ’F OR (C AND D) OR

 

 

G’,1"

 

 

integer from 1 to <number_of_levels>

<N>

<to_level_

integer from 1 to <number_of_levels>

number>

 

<number_of_

integer from 1 to the number of existing sequence levels (maximum 10)

levels>

 

<branch_

<qualifier> see "Qualifier" on page 12-6

qualifier>

 

 

 

 

Example

OUTPUT XXX;":MACHINE1:TTRIGGER:BRANCH1 ’ANYSTATE’, 3"

 

 

OUTPUT XXX;":MACHINE2:TTRIGGER:BRANCH2 ’A’, 7"

 

 

OUTPUT XXX;":MACHINE1:TTRIGGER:BRANCH3 ’((A OR B) OR NOTG)’,

 

 

1"

 

 

 

12–10