SFORmat Subsystem
MASTerMASTer
Command:MACHine{12}:SFORmat:MASTer <clock_id>, <clock_spec>
<clock_id>
<clock_spec>
Example
Query
Returned Format
Example
The MASTer clock command allows you to specify a master clock for a given machine. The master clock is used in all clocking modes (Master, Slave, and Demultiplexed). Each command deals with only one clock (J,K,L,M); therefore, a complete clock specification requires four commands, one for each clock. Edge specifications (RISing, FALLing, or BOTH) are ORed.
At least one clock edge must be specified.
{JKLM}
{OFFRISingFALLingBOTH}
OUTPUT XXX;":MACHINE2:SFORMAT:MASTER J, RISING"
:MACHine{12}:SFORmat:MASTer? <clock_id>
The MASTer query returns the clock specification for the specified clock.
[:MACHine{12}:SFORmat:MASTer] <clock_id>,<clock_spec><NL>
OUTPUT XXX;":MACHINE2:SFORMAT:MASTER? <clock_id>"