SLISt Subsystem

OPATtern

Query:MACHine{12}:SLISt:MMODe?

The MMODe query returns the current marker mode selected.

Returned Format [:MACHine{12}:SLISt:MMODe] <marker_mode><NL>

Example

OUTPUT XXX;":MACHINE1:SLIST:MMODE?"

 

 

OPATtern

Command:MACHine{12}:SLISt:OPATtern <label_name>,<label_pattern>

The OPATtern command allows you to construct a pattern recognizer term for the O Marker which is then used with the OSEarch criteria when moving the marker on patterns. Because this command deals with only one label at a time, a complete specification could require several invocations.

When the value of a pattern is expressed in binary, it represents the bit values for the label inside the pattern recognizer term. In whatever base is used, the value must be between 0 and 232 1, since a label may not have more than 32 bits. Because the <label_pattern> parameter may contain don’t cares, it is handled as a string of characters rather than a number.

<label_name> string of up to 6 alphanumeric characters

<label_pattern> "{#B{01X} . . . #Q{01234567X} . . . #H{0123456789ABCDEFX} . . . {0123456789} . . . }"

Example

OUTPUT

XXX;":MACHINE1:SLIST:OPATTERN

’DATA’,’255’ "

 

OUTPUT

XXX;":MACHINE2:SLIST:OPATTERN

’ABC’,’#BXXXX1101’ "

 

 

 

 

7–11