STRigger (STRace) Subsystem

TIMER

Query

:MACHine{12}:STRigger:TERM? <term_id>,

 

<label_name>

The TERM query returns the specification of the term specified by term identification and label name.

Returned Format [:MACHine{12}:STRAce:TERM] <term_id>,<label_name>,<pattern><NL>

Example

OUTPUT XXX;":MACHINE1:STRIGGER:TERM? B,’DATA’ "

 

 

TIMER

Command:MACHine{12}:STRigger:TIMER{12} <time_value>

<time_value>

Example

Query

Returned Format

Example

The TIMER command sets the time value for the specified timer. The limits of the timer are 400 ns to 500 seconds in 16 ns to 500 μs increments. The increment value varies with the time value of the specified timer. There are two timers and they are available for either machine but not both machines simultaneously.

real number from 400 ns to 500 seconds in increments which vary from 16 ns to 500 μs.

OUTPUT XXX;":MACHINE1:STRIGGER:TIMER1 100E6"

:MACHine{12}:STRigger:TIMER{12}?

The TIMER query returns the current time value for the specified timer.

[:MACHine{12}:STRigger:TIMER{12}] <time_value><NL>

OUTPUT XXX;":MACHINE1:STRIGGER:TIMER1?"

6–21

Page 109
Image 109