TWAVeform Subsystem
MLENgth
|
| Inserting Oscilloscope Waveforms |
Command | :MACHine{12}:TWAVeform:INSert <module_spec>, | |
|
| <label_name> |
|
| This inserts a waveform from an oscilloscope to the timing waveforms display. |
<module_spec> | {12345678910} slot in which the oscilloscope master card is | |
|
| installed |
<label_name> | string of one alpha and one numeric character | |
|
|
|
Example | OUTPUT XXX;":MACHINE1:TWAVEFORM:INSERT 3, ’C1’" | |
|
|
|
MLENgth
Command:MACHine{12}:TWAVeform:MLENgth <memory_length>
The MLENgth command allows you to specify the analyzer memory depth. Valid memory depths range from 4096 samples through the maximum system memory depth minus 8192 samples (HP 16554A/HP 16555A) or the maximum system memory depth minus 16384 (HP 16555D). Memory depth is affected by acquisition mode. If the <memory_depth> value sent with the command is not a legal value, the closest legal setting will be used.
<memory_length> {4096 8192 16384 32768 65536 131072 262144 516096 524288 1040384 (HP 16554A)
524288 1040384 1048576 2088960 (HP 16555A) 524288 1048576 2080768 2088960 4177920 (HP 16555D)}
Example | OUTPUT XXX;":MACHINE1:TWAVEFORM:MLENGTH 262144" |
|
|