WLISt Subsystem

LINE

Inserting Oscilloscope Waveforms

Command:WLISt:INSert <module_spec>,<label_name>

This inserts a waveform from an oscilloscope to the timing waveforms display.

<module_spec> {12345678910} slot in which master card is installed

<label_name> string of one alpha and one numeric character, identical to that on the oscilloscope waveform display.

Example

OUTPUT XXX;":WLIST:INSERT 3, ’C1’"

 

 

Command

<line_num_mid_ screen>

Example

LINE

:WLISt:LINE <line_num_mid_screen>

The LINE command allows you to scroll the timing analyzer listing vertically. The command specifies the state line number relative to the trigger. The analyzer then highlights the specified line at the center of the screen.

integer from -516096 to +516096 (HP 16554A) or -1040384 to +1040384 (HP 16555A)

or -2080768 to +2080768 (HP 16555D).

OUTPUT XXX;":WLIST:LINE 0"

4–7

Page 63
Image 63
HP 16555D, 16555A, 16554A manual Line, CommandWLIStINSert modulespec,labelname, WLIStLINE linenummidscreen