SFORmat Subsystem

MOPQual

MOPQual

Command:MACHine{12}:SFORmat:MOPQual <clock_pair_id>, <qual_operation>

<clock_pair_ id>

<qual_ operation>

Example

Query

Returned Format

Example

The MOPQual (master operation qualifier) command allows you to specify either the AND or the OR operation between master clock qualifier pair 1/2, or between master clock qualifier pair 3/4. For example, you can specify a master clock operation qualifier 1 AND 2.

{12} where 2 indicates qualifier pair 3/4.

{ANDOR}

OUTPUT XXX;":MACHINE1:SFORMAT:MOPQUAL 1,AND"

:MACHine{12}:SFORmat:MOPQual? <clock_pair_id>

The MOPQual query returns the operation qualifier specified for the master clock.

[:MACHine{12}:SFORmat:MOPQUal <clock_pair_id>] <qual_operation><NL>

OUTPUT XXX;":MACHine1:SFORMAT:MOPQUAL? 1"

5–11