DATA and SETup Commands
Data Preamble Description
Byte Position
374 bytes - List of pods in this analyzer, where a binary 1 indicates that the corresponding pod is assigned to this analyzer
bit 31 | bit 30 | bit 29 | bit 28 | bit 27 | bit 26 | bit 25 | bit 24 |
unused | unused | unused | unused | unused | unused | unused | unused |
bit 23 | bit 22 | bit 21 | bit 20 | bit 19 | bit 18 | bit 17 | bit 16 |
|
|
|
|
|
|
|
|
unused | unused | clkpd1 | unused | unused | unused | unused | unused |
bit 15 | bit 14 | bit 13 | bit 12 | bit 11 | bit 10 | bit 9 | bit 8 |
|
|
|
|
|
|
|
|
unused | unused | unused | Pod 12 | Pod 11 | Pod 10 | Pod 9 | Pod 8 |
bit 7 | bit 6 | bit 5 | bit 4 | bit 3 | bit 2 | bit 1 | bit 0 |
|
|
|
|
|
|
|
|
Pod 7 | Pod 6 | Pod 5 | Pod 4 | Pod 3 | Pod 2 | Pod 1 | unused |
Example
41
45
49
53
Example
xxxxxxxx xx1x xxxx xxx0 0000 0001 111x indicates that data pods 1 through 4 and clock pod 1 are assigned to this analyzer (x = unused).
4 bytes - Master chip for this analyzer.
4 bytes - Maximum hardware memory depth available for this analyzer. 4 bytes - Unused.
8 bytes - Sample period in picoseconds (timing only).
The following 64 bits represent a sample period of 8,000 picoseconds (8 nanoseconds):
00000000 00000000 00000000 00000000 00000000 00000000 00011111 01000000