Module Level Commands

ARMLine

ARMLine

Command:ARMLine MACHine<N>

The ARMLine command selects which machine generates the arm out signal on the IMB (intermodule bus). This command is only valid when two analyzers are on. However, the query is always valid.

<N> {12}

Example

OUTPUT XXX;":ARMLINE MACHINE1"

 

 

Query:ARMLine?

If the analyzer is set up for OR’d triggering, then the ARMLine query returns an empty string. This cannot be used for setting up OR’d triggering.

Returned Format [:ARMLine]{MACHine<N>}<NL>

Example

OUTPUT XXX;":ARMLine?"

 

 

DBLock

Command:DBLock {PACKed UNPacked}

The DBLock command specifies the data block format that is contained in the response from a :SYSTem:DATA? query. See Chapter 17 for more information on the :SYSTem:DATA command and query.

The PACKed option (default) uploads data in a compressed format. This option is used to upload data for archiving, or for reloading back into the analyzer. When an analyzer configuration is saved to disk, the PACKed data format is always used (regardless of the current DBLock selection).

2–5

Page 39
Image 39
HP 16555D, 16555A, 16554A manual CommandARMLine MACHineN, QueryARMLine?, CommandDBLock PACKed UNPacked