Table B-1. Actions and Warnings

 

Code, Message

Cause

What to Do

 

 

 

W-VS-2 - Vacuum Pressure Too High

The vacuum pressure of the heads is

Call your Authorized Service Provider.

 

too high.

 

W-VS-3 Vacuum Pressure Not Recov- ering

The vacuum pressure of the heads is not recovering after an air purge.

Check vacuum pressure system to ensure it is working properly and check vacuum tubing for leaks.