Calibration
Selecting a Calibration Starting Point
When printing a calibration target (see page 79), you must select a calibration starting point. Each starting point serves a different purpose. Use the starting point that best suits your needs.
✔Current settings: Starts the calibration using the current calibration data. You will use Current settings most often. Use Current settings if the printer is close to optimal and just needs some
✔Factory defaults: Starts the calibration using the settings that were shipped with the printer. Use this option if calibrating the printer for the first time or if the current calibration is unacceptable and you would like to make a fresh start.
✔Redo last calibration: Starts the calibration from the beginning of the last calibration. Use this option if the calibration was in tolerance or very nearly in tolerance up until the most recent calibration.
✔Designated event: Starts the calibration at a specific calibration event. Use this option if you know the date and time when the last good calibration was done and you want to start the calibration from those settings. If you select this option, an extra dialog box in the Print Target Wizard requires you to select the event from which to start the calibration.
| Contents |
| Back |
| Next |
| Index | 78 | |
|
|
|
|
|
|
|
|
|