Calibration
✔For analysis on a strip reading densitometer
a.Prepare the target for reading in the densitometer.
b.Click Next.
✔For file analysis
a.Read the calibration target and save the results to a file.
b.Click Next.
c.Search for and select the file in which you saved the target readings.
d.Click OK.
5.Based on the analysis, indicate what to do with the calibration data. You must consider whether the calibration is in tolerance, and which setting you used for a starting point to determine the proper action.
✔Record: record the data and consider this calibration complete. This results in the calculation of a new set of LUTs. You may want to use this option if the printer is only slightly out of tolerance and the current production appears normal OR if the starting point was other than Current Settings.
✔Discard: discard the calibration data. The Next button changes to a Finish button. You should only select this option if the printer is clearly in tolerance, when the starting point was Current Settings, and when you do not want to disturb the current printer settings for the media.