71
C4 Manual Rev 7.5.2
7.6.4 SL Set UUT Temperature Limits
The SL command sets the lower UUT limit (setup parameter F27) and the upper
UUT limit (setup parameter F28). For a detailed explanation of the effect of these
parameters, see Sections 3.5 and 4. The command sets the lower and upper
temperature limits at the same time. The syntax of the command is:
SL lll.l uuu.u<CR><LF>
where: lll.l is the lower limit (F27)
uuu.u is the upper limit (F28)
lll.l & uuu.u are one decimal place numbers
Example:
SL -75 130<CR><LF>
7.6.5 SD Set UUT Temperature Differential Limits
The SD command sets the lower UUT temperature differential limit (setup
parameter F29) and the upper UUT temperature differential limit (setup
parameter F30). For a detailed explanation of the effect of these parameters, see
Section 4. The command sets the lower and upper temperature differential limits
at the same time. The syntax of the command is:
SD lll.l uuu.u<CR><LF>
where lll.l is the lower differential limit (F29)
uuu.u is the upper differential limit (F30)
lll.l & uuu.u are one decimal place floating point numbers
Example:
SD 50 30<CR><LF>
7.6.6 UP Write Current Parameters to NV Memory
UP writes the current SRAM resident setup parameters to EEPROM non-volatile
setup parameter tables. See Section 7.6 for explanation. This command uses no
arguments. The syntax of the command is:
UP<CR><LF>