7.6.4 SL Set UUT Temperature Limits

The SL command sets the lower UUT limit (setup parameter F27) and the upper UUT limit (setup parameter F28). For a detailed explanation of the effect of these parameters, see Sections 3.5 and 4. The command sets the lower and upper temperature limits at the same time. The syntax of the command is:

SL lll.l uuu.u<CR><LF>

where:

lll.l is the lower limit (F27)

 

uuu.u is the upper limit (F28)

 

lll.l & uuu.u are one decimal place numbers

Example:

SL -75 130<CR><LF>

7.6.5 SD Set UUT Temperature Differential Limits

The SD command sets the lower UUT temperature differential limit (setup parameter F29) and the upper UUT temperature differential limit (setup parameter F30). For a detailed explanation of the effect of these parameters, see Section 4. The command sets the lower and upper temperature differential limits at the same time. The syntax of the command is:

SD lll.l uuu.u<CR><LF>

where lll.l is the lower differential limit (F29) uuu.u is the upper differential limit (F30)

lll.l & uuu.u are one decimal place floating point numbers

Example:

SD 50 30<CR><LF>

7.6.6 UP Write Current Parameters to NV Memory

UP writes the current SRAM resident setup parameters to EEPROM non-volatile setup parameter tables. See Section 7.6 for explanation. This command uses no arguments. The syntax of the command is:

UP<CR><LF>

C4 Manual Rev 7.5.2

71

Page 71
Image 71
Sigma C4 manual Upcrlf