Step

Displayed

 

 

Step .

Keys to press

Description of entry

Number

Substep

 

 

 

 

 

 

00.0

CLEAR/ENTRY

45.2

ENTER

Go to 45.2/C

 

00.1

CLEAR/ENTRY

5

ENTER

Use 5 minute ramp time

0

00.2

CLEAR/ENTRY

2

ENTER

Hold temperature for 2 minutes

 

00.3

CLEAR/ENTRY

1

ENTER

Execute step 1 next

 

00.4

CLEAR/ENTRY

1

ENTER

Use probe 1

 

01.0

CLEAR/ENTRY

32.3

ENTER

Go to 32.3/C

 

01.1

CLEAR/ENTRY

2

ENTER

Use 2 minute ramp time

1

01.2

CLEAR/ENTRY

6

ENTER

Hold 32.3/C for 6 minutes

 

01.3

CLEAR/ENTRY

2

ENTER

Execute step 2 next

 

01.4

CLEAR/ENTRY

1

ENTER

Use probe 1

 

02.0

CLEAR/ENTRY

50.1

ENTER

Go to 50.1/C

 

02.1

CLEAR/ENTRY

0

ENTER

Ramp as quickly as possible

2

02.2

CLEAR/ENTRY

9900

ENTER

Do not hold at this temperature

 

02.3

CLEAR/ENTRY

3

ENTER

Execute step 3 next

 

02.4

CLEAR/ENTRY

1

ENTER

Use probe 1

 

03.0

CLEAR/ENTRY -10.0

Go to -10/C

 

03.1

CLEAR/ENTRY

0

ENTER

Ramp as quickly as possible

3

03.2

CLEAR/ENTRY

3

ENTER

Hold -10/C for 3 minutes

 

03.3

CLEAR/ENTRY

100

ENTER

End program

 

03.4

CLEAR/ENTRY

1

ENTER

Use probe 1

90

C4 Manual Rev 7.5.2

Page 90
Image 90
Sigma C4 manual 00.0, 00.2, 00.3, 00.4, 02.0, 02.2, 02.3, 02.4, 03.0, 03.2, 03.3, 03.4