8. SETUP MODE

Setup mode is entered by placing the mode switch in the right most position. Setup mode is used to modify operational parameters of the chamber. When first entering Setup mode, the digital display should display:

f-0n (n = 1 to 10)

This first display indicates the setting for field 0. There are thirty parameter fields which can be modified. Table 8.1 shows the meaning of each field and the values each field can be assigned.

 

 

 

 

 

 

T A B L E

8.1

 

Field

 

Field Purpose

 

Permitted Field

Factory Setting

#

 

 

Values

 

 

 

 

 

 

 

 

 

 

 

 

PID controller, Proportional term

 

4 - 7

0

Controller gain, P term, number=°C from

1 - 10

Default=4

 

setpoint where controller will use full on

 

 

 

 

control

 

 

 

 

1

Number of Probes installed

1 - 2

1

 

 

 

 

2

Remote communication port select

0 - 1

1

0 = EIA-232

1 = IEEE-488

 

 

 

 

EIA-232 port baud rate

 

 

3

1=300

2=600

3=1200

4=2400

1 - 8

6

5=4800

6=9600

7=19,200

 

 

 

 

 

 

8=38,400

 

 

 

 

 

 

 

4

EIA-232 port data bits

7 - 8

8

 

 

 

 

5

EIA-232 port stop bits

1 - 2

1

 

 

 

 

6

EIA-232 port parity enable

0 - 1

0

0=disabled

1=enabled

 

 

 

7

EIA-232 port parity type

0 - 1

0

 

0=even

1=odd

 

 

 

 

 

 

8

 

EIA-232 port echo

0 - 1

1

0=disabled

1=enabled

 

 

 

9

IEEE-488 port primary address

0 - 30

1