Step

Displayed

 

 

 

 

Step .

Keys to press

Description of entry

Number

Substep

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

02.0

CLEAR/ENTRY

11.5

ENTER

Go to 11.5/C

 

02.1

CLEAR/ENTRY

15

ENTER

Use 15 minute ramp time

2

02.2

CLEAR/ENTRY

20

ENTER

Hold temperature for 20 minutes

 

02.3

DISP/CNTL

 

Accept default value of 3

 

 

Execute step 3 next

 

02.4

CLEAR/ENTRY

2

ENTER

Use probe 2

 

03.0

CLEAR/ENTRY

-32.3 ENTER

Go to 32.3/C

 

03.1

CLEAR/ENTRY

100

ENTER

Use 1 hour ramp time

3

03.2

CLEAR/ENTRY

100

ENTER

Hold -32.3/C for 1 hour

 

03.3

DISP/CNTL

 

Accept default value of 4

 

03.4

CLEAR/ENTRY 2

ENTER

Use probe 2

 

 

 

04.0

DISP/CNTL

 

Go to 50.1/C

 

04.1

DISP/CNTL

 

Accept default value of 0

4

04.2

CLEAR/ENTRY

9900

ENTER

Do not hold at this temperature

 

04.3

DISP/CNTL

 

Accept default value of 3

 

04.4

CLEAR/ENTRY

5

ENTER

Turn compressor port off

 

05.0

CLEAR/ENTRY

50.0 ENTER

Go to 50/C

 

05.1

DISP/CNTL

 

Accept default value of 0

5

05.2

CLEAR/ENTRY

230

ENTER

Hold 50/C for 2 hours & 30

 

05.3

DISP/CNTL

 

Accept default value of 6

 

 

Execute step 6 next

 

05.4

CLEAR/ENTRY

2

ENTER

Use probe 2

 

06.0

CLEAR/ENTRY 5

ENTER

Set loop count to 5

 

06.1

DISP/CNTL

 

Accept default value

6

06.2

DISP/CNTL

 

Accept default value

 

06.3

DISP/CNTL 1 ENTER

Step number of first step in loop

 

06.4

CLEAR/ENTRY

3

ENTER

Loop command

94

C4 Manual Rev 7.5.2

Page 94
Image 94