C4 Manual Rev 7.5.2
1.5.18 Remote Mode EIA-232 Port Initialization
The CC-3 required that to use the EIA-232 port, the port had to be initialized by
switching the mode switch to SETUP before switching to REMOTE mode. The C4
eliminates this requirement. The EIA-232 port is initialized each time the remote
switch is switched to REMOTE mode.
1.5.19 Remote Mode Fault Tolerant Parser
The C4 uses a very fault tolerant parser. Command strings received over either
the EIA-232 or GPIB ports are converted to upper case, extra spaces and tabs are
removed, commas are converted to spaces and line terminators are corrected if
necessary. For this reason, programs that work properly with the C4 and not with
a CC-3, that use only CC-3 commands, probably have syntax errors that the C4
parser corrects. See Section 9.7.7
1.5.20 Remote Mode System Information Queries
QV Query Firmware Version ..................... See Section 7.4.1
QN Query Serial Number ........................ See Section 7.4.2
QR Query Temperature Range.................... See Section 7.4.3
QS Query Setpoint & Control Probe Number ........ See Section 7.4.4
QF Query Setup Parameter Data (Binary Query)..... See Section 7.4.5
QFA Query Setup Parameter Data (ASCII Query) ..... See Section 7.4.5
1.5.21 Remote Mode Operation Information Queries
RSA Request Status Byte (ASCII Query) ............. See Section 7.5.1
REA Request Error Byte (ASCII Query) ............. See Section 7.5.2
QE Query Error/Status String (Binary Query) ....... See Section 7.5.3
QEA Query Error/Status String (ASCII Query) ........ See Section 7.5.3
QC Query Last Command........................ See Section 7.5.4