Field
#Field Purpose Permitted Field
Values Factory Setting
82 C4 Manual Rev 7.5.2
10
PID controller, Integral term
Tens of seconds required for I term effect
equal to P term effect
Adjustment not
normally
required
5 - 7
Default = 5
11
PID controller, Differential term
Seconds required for D term effect to equal
P term effect
Adjustment not
normally
required
5 - 7
Default = 5
12 PID controller, integral wind-up
limiter term
Adjustment not
required 7
13
Refrigeration compressor front
panel control
1=yes 0=no
0 - 1 0
14 Autostart in Local Mode
1=yes 0=no 0 - 1 0
15 Blower shut-off mode
0=disable 1=enable 0 - 1 0
16 Temperature Scale
0=Celsius 1=Fahrenheit 0 - 1 0
17 Probe 1 Correction
Uncorrected first value (U1) See Section 8.7 0
18 Probe 1 Correction
Corrected first value (C1) See Section 8.7 0
19 Probe 1 Correction
Uncorrected second value (U2) See Section 8.7 100
20 Probe 1 Correction
Corrected second value (C2) See Section 8.7 100
21 Probe 2 Correction
Uncorrected first value (U1) See Section 8.7 0
22 Probe 2 Correction
Corrected first value (C1) See Section 8.7 0
23 Probe 2 Correction
Uncorrected second value (U2) See Section 8.7 100
24 Probe 2 Correction
Corrected second value (C2) See Section 8.7 100