No Probe 2 Error
If a program step calls for control by probe 2 when setup parameter F1 = 1 (number of probes) the C4 will generate a No Probe 2 Error. Example
007 nop2
You must either remove the step that calls for probe 2 or enable probe 2 with setup parameter F1 (See Section 8) before the program will run. See Section 8.3.
Invalid Loop Count Error
If a program step is a loop start step (substep 4, probe value = 3), then the loop count (substep 0, loop count) must be in the range 1 - 999 integer (no decimal value). If the number is either not an integer, or is outside the range 1 - 999, the
C4 will generate an Invalid Loop Count Error. Example:
003 loop
The invalid loop count must be corrected before the program will run. See Section 6.8.1.
Setpoint Out of Range Error
If a program step requests a setpoint outside the System Operating Range (See Section 3.5), the C4 will generate a Setpoint Out of Range Error. The error message will be preceded by a minus sign
005 sor | 005 |
Either the setpoint request or one of the temperature ranges that defines the System Operating Range will have to be changed before the program will run.
52 | C4 Manual Rev 7.5.2 |