92 C4 Manual Rev 7.5.2
Step
Number
Displayed
Step .
Substep
Keys to press Description of entry
0
00.0 CLEAR/ENTRY 45.2 ENTER Go to 45.2/C
00.1 CLEAR/ENTRY 5 ENTER Use 5 minute ramp time
00.2 CLEAR/ENTRY 2 ENTER Hold temperature for 2 minutes
00.3 DISP/CNTL Accept default value of 1
00.4 DISP/CNTL Accept default value of 1
1
01.0 CLEAR/ENTRY 32.3 ENTER Go to 32.3/C
01.1 CLEAR/ENTRY 2 ENTER Use 2 minute ramp time
01.2 CLEAR/ENTRY 6 ENTER Hold 32.3/C for 6 minutes
01.3 DISP/CNTL Accept default value of 2
01.4 DISP/CNTL Accept default value of 1
2
02.0 CLEAR/ENTRY 50.1 ENTER Go to 50.1/C
02.1 DISP/CNTL Accept default value of 0
02.2 CLEAR/ENTRY 9900 ENTER Do not hold at this temperature
02.3 DISP/CNTL Accept default value of 3
02.4 DISP/CNTL Accept default value of 1
3
03.0 CLEAR/ENTRY -10.0 ENTER Go to -10/C
03.1 DISP/CNTL Accept default value of 0
03.2 CLEAR/ENTRY 3 ENTER Hold -10/C for 3 minutes
03.3 CLEAR/ENTRY 100 ENTER End program
03.4 DISP/CNTL Accept default value of 1
To run this program, set the program pointer to the first program step (00) and
then start the program using the following key sequence:
00 <DISP/CNTL><START/STOP>