parameter F15. Likewise, at the end of a local or remotely controlled program, chamber blowers will continue to run, but heating and cooling will be disabled. Note that a chamber in this condition, with blowers running, will exhibit some heating due to blower air friction. This effect is exaggerated in units equipped with high velocity blowers. Moving the rotary switch between OFF and LOCAL will maintain the last used setpoint for the next operation of the controller. See 6.3 for description of auto-start function.

2.1.4 Keyboard

The keyboard consists of 20 momentary contact keys. Certain keys are functional only in some of the modes of operation. With some keys, such as the Display Control, the function is different depending on the mode of operation. The Keyboard is arranged in the following configuration:

1

2

3

CLEAR

START

PROG

STOP

 

 

 

 

 

4

5

6

REFER

ENTER

 

 

 

 

 

7

8

9

ADV

DISP

TEMP

 

 

 

 

 

.

0

+

CLEAR

DISP

-

ENTRY

CNTL

 

 

 

 

 

CLEAR PROG

Clear program from memory

START STOP

Start/Stop temperature control

REFER

Toggle refrigeration compressor on or off *

ENTER

Enter/finalize current keyboard entry

ADV

Advance to next program step

DISP TEMP

Display temperature

CLEAR ENTRY

Clear current keyboard/display entry

DISP CNTL

Display control setpoint

* = LOCAL MODE ONLY.

C4 Manual Rev 7.5.2

21

Page 21
Image 21
Sigma C4 manual Clear Prog