CHAPTER 7 RS232 & RS485 REMOTE CONTROL
7.1 INTRODUCTION
This Chapter describes the operation of the GenesysTM 3300W power supplies via the serial communication port. Details of the initial
7.2 CONFIGURATION
7.2.1 Default setting
The power supply is shipped with the following settings:
6 | Off | ||
9600 | |||
RS232 |
| Maximum | |
0 | 0 | ||
Maximum | Off | ||
H1 (Master) | Unlocked (UFP) |
7.2.2Address setting
The power supply address can be set to any address between 0 and 30. Follow the instructions described below to set the unit address.
1.If the unit is in Remote mode (front panel REM/LOC LED illuminated), press the REM/LOC button to put the unit into Local mode.
2.Press and hold for the REM/LOC button for approximately 3 sec. The VOLTAGE display will indicate the unit address.
3.Using the VOLTAGE adjust encoder, select the unit address.
To preview the power supply address at any time, press and hold the REM/LOC button for approx. 3 sec. The VOLTAGE display will indicate the power supply address.
7.2.3RS232 or RS485 selection
To select between RS232 or RS485 set the rear panel setup switch
7.2.4Baud Rate setting
Five optional Baud rates are possible: 1200, 2400, 4800, 9600 and 19200. To select the desired rate, the following steps should be taken:
1.If the unit is in Remote mode (front panel REM/LOC LED illuminates), press REM/LOC button to put the unit into Local mode.
2.Press and hold the REM/LOC button for approximately 3 sec. The CURRENT display will show the unit Baud Rate.
3.Using the CURRENT adjust encoder, select the desired Baud Rate.
7.2.5Setting the unit into Remote or Local mode
1. The unit will be put into Remote mode only via serial communication command. Commands that will put the unit into Remote mode are:
RST | PV n |
OUT n | PC n |
RMT n |
|
(for n values see Tables
50