83-507-5002 Rev. B
58
25 RCL Recalls last settings. Settings are from the last power-down or from the last
“SAV” command.
26 MODE? Returns the power supply operation mode. When the power supply is On
(OUT 1) it will return “CV” or “CC”. When the power supply is OFF (OUT 0 or
fault shutdown) it will return “OFF”.
27 MS? Returns t he Master/Slave setting. Master: n= 1, 2, 3, or 4 Slave: n=0
NOTES:
1. In Advanced parallel mode (refer to Sec. 5.15.2), “n” is the total system current.
2. In Advanced parallel mode, “MC?” returns the Master unit current multiplied by t he number of
slave units +1.
7.9 GLOBAL OUTPUT COMMANDS
7.9.1 GENERAL
All supplies, even if not the currently addressed supply, receiving a global command will execute
the command. No response to the PC issuing the command will be returned to the PC. The PC
issuing the command will be responsible to delay and any other communications until the com-
mand is execute. 200 Ms minimum is the suggested delay.
If the command contains an error, out of range values for example, no error report will be sent to
the issuing PC.
Table 7-3
GRST Reset. Brings the power supply to a safe and known state:
Output voltage: 0V, output current: 0A, OUT: Off, Remote: RMT 1,
AST: Off OVP: Max, UVL: 0.
The conditional register (FLT and STAT) are updated. Other registers are not
changed.
Non-Latching faults (FB, OVP, SO) are cleared, OUT fault stays
GPV n Sets the output voltage value in volts. The range of voltage values is shown in
Table 7-5. ‘n’ may be up to 12 char plus dec. pt
GPC n Program the output current value in amperes. T he range of current values is
shown in Table 7-6. ‘n’ may be up to 12 char plus dec. pt
GOUT Turns the output to ON or OFF:
“OUT 1/ON” = turn on
“OUT 0/OFF” = turnoff, clears CV and CC bits in the Status Condition (STAT).
OUT ON will respond with “E07’ if the output cannot be turned on because of a
latching fault (OTP< AC, ENA, SO) shutdown.
GSAV Save present settings. Same settings as power-down last settings listed in Er-
ror!
Reference source not found. Except the address and Baud rate are not saved
Saves to the RAM. These settings are erased when the supply power is
switched off and the new ‘last settings’ are saved.
GRCL Recall last settings. Settings are from last power-
down or from last ‘SAV’ or
‘GSAV’ command. Address and Baud rate are not recalled so comm unication is
not interrupted.