AMEasure Subsystem (Audio Analyzer)
AM700 Audio Measurement Set Programmer Manual 3–19
AMEasure:SETTled:PDIFference:RESolution <numeric_value>
Usage: Sets or queries the global settling resolution setting of the phase difference for
settled data points.
Range: 0 to 180 degrees.
Query: AME:SETT:PDIF:RES? returns the global setting for phase difference
resolution for settled data points.
*RST: 1.000 degree.
AMEasure:SETTled:PDIFference:TOLerance <numeric_value>
Usage: Sets or queries the global settling tolerance setting of the phase difference
measurement.
Range: 0 to 100%
Query: AME[1–4]:SETT:PDIF:TOL? returns the global setting for phase difference
tolerance for settled data points.
*RST: 1.000%
AMEasure:SETTled:SEParation:RESolution <numeric_value>
Usage: Sets or queries the global settling resolution setting of the separation measure-
ment.
Range: 0 to 100%
Query: AME:SETT:SEP:RES? returns the global setting for resolution for settled
data points.